Enjoy FREE SHIPPING on all FreeStyle Libre 2 purchases until 31 May 2024. Simply enter promo code FREESHIP at checkout and click ‘SUBMIT CODE’ to apply!

Pharmac proposes to commence FreeStyle Libre 2 subsidy for people living with type 1 diabetes! Read more

Order pick-ups are not available

X
Menu
Cart
X

Search Website

QUICK ORDER PRODUCTS

If you know the product code enter it here  

Advanced Dissolved Oxygen Meter

HA-HI6421-02

HA-HI6421-02 is a streamlined dissolved oxygen benchtop meter supplied with Hanna’s optical dissolved oxygen probe (opdo®) and features a large touch screen display for fresh and saltwater measurements

Brand: Hanna

Learn more

Product CodeProduct Product Description Price (Excl gst) Qty Add
HA-HI6421-02 Advanced Dissolved Oxygen Meter Advanced Dissolved Oxygen Meter Login for price  

Go to checkout

HA-HI6421-02 is a streamlined benchtop meter with a large touch screen display, comprised of a
housing and an integrated module designed for fresh and saltwater measurements of
dissolved oxygen.
HA-HI6421-02 includes Hanna's HI7641133 optical dissolved oxygen probe (opdo®) that is based on
the principle of fluorescence quenching. An immobilized Pt-based luminophore is excited by
the light of a blue LED and emits a red light. As oxygen interacts with the luminophore it
reduces the intensity and lifetime of the luminescence. The lifetime of the luminescence is
measured by a photodetector and is used to calculate the dissolved oxygen concentration.
The probe is fitted with easy to use Smart Caps (HI764113-1) which lock in place and contain
pre-loaded calibration coefficients that are automatically transmitted to the probe. The Smart
Cap features an immobilized O₂ sensitive luminophore with rugged insoluble black oxygen
permeable protective layer.
Over time, the sensor’s optical components can age but are compensated for by using the
reference signal to compensate the measuring path. As a result, the sensor provides accurate
DO measurements over long periods of time without the need for frequent calibration.
Concentration measured.
Additional features include built-in methods and calculations for the measurement of BOD
(Biological Oxygen Demand), OUR (Oxygen Uptake Rate), and SOUR (Specific Oxygen Update
Rate).
Pressure compensation is done automatically (built-in barometer) or users have the option to
manually enter required value. Pressure is displayed in user-configurable units: mmHg, mbar,
kPa, inHg, psi, atm.
• User interface
o 7-inch capacitive touch screen with multi-touch support

o Capacitive touch back, home and system menu keys
o User-friendly icons and symbols allow users to easily navigate and interpret
the instrument functions.
• The user can select between five different views:
o Basic measurement configuration
o Simple GLP with calibration information
o Full GLP with electrode status and calibration point details
o Live updated, interactive graph
o Tabulated data with date, time, and notes
• Measurement
o Measure %Sat, mg/L, ppm (DO); mg/L, ppm (BOD); mg/L (OUR); ppm, mg/L
(SOUR)
o Application-specific profiles allow quick and direct measurement without the
need to update the sensor and system settings
o Active log during measurement
o Measurement stability indicator (using the Stability Criteria setting)
o Reading modes: direct and direct/autohold
o Temperature compensation can be Automatic or set manually
o Audible and/or alarm messages for measurements outside of predefined limits
o Galvanic isolation for measurement
• Calibration
o One or two points calibration at 0% and/or 100% saturation
o Single point manual calibration in mg/L or % saturation using a reference
method
o Non-volatile memory saves data and settings
• Logging
o Data log collection of at least 1,000,000 data points (with time and date
stamp)
o Logging types: manual, automatic, autohold
o Sample ID for manual and Autohold data
• Connectivity features & services
o Transfer logged data to a USB thumb drive
o Log files that include measurements and calibration data (as .csv file)
o FTP and email for log export via Ethernet and Wi-Fi connection
o USB type A for USB stick, keyboard, and printer
o USB type C for USB stick and PC connection
• Help section for meter guidance
o Video support presentation of main functionalities

DO (HI7641133
Optical probe)
Range 0.00 to 50.00 mg/L (ppm) concentration; 0.0 to
500.0 % saturation
Resolution 0.01 mg/L (ppm); 0.1 % saturation
Accuracy from 0.00 to 20.00 mg/L (ppm) 1.5 % of reading or
± 0.01 mg/L (ppm), whichever is greater
from 20.00 to 50.00 mg/L (ppm) ±5 % of reading from 0.0 to 200.0 % saturation ± 1.5 % of reading or ± 0.1 %, whichever is greater
from 200.0 to 500.0 % saturation ±5 % of reading
Barometric Pressure Range 450 to 850 mmHg; 600 to 1133 mBar; 60 to 133 kPa;
17 to 33 inHg; 8.7 to 16.4 psi; 0.592 to 1.118 atm
Resolution 1 mmHg; 1 mBar; 1 kPa; 1 inHg; 0.1 psi; 0.001 atm
Accuracy ±3 mmHg within ±15 % from the calibration point
±3 mmHg ±1 least signicant digit
Temperature Range 20.0 to 120.0 °C
4.0 to 248.0 °F
253.0 to 393.0 K
Resolution 0.1 °C; 0.1 °F; 0.1 K
Accuracy ±0.2 °C; ±0.4 °F; ±0.2 K
DO Calibration Points One or two points automatic calibration at 100% (8.26 mg/L) and 0% (0 mg/L).
Single point manual using a value entered by the
user in % saturation or mg/L.
Standards 0 and 100% saturation
  Reminder Disabled
Daily: 0 min. to 23 hours and 59 min.
Periodic: 1 min. to 500 days, 23 hours and 59 min.
Temperature Compensation Automatic or Manual
Salinity Compensation Automatic from 0 to 70 PSU (manually set)
0.0 to 70.0 % / 0.0 to 45.0 g/L / 0.0 to 42.0 psu
Reading Modes Direct
Direct/Autohold
Stability criteria Accurate Medium Fast
Isopotential 7.000 or 4.010
Sampling rate 1000 ms
DO Views Basic Measurement (DO, Temperature) Stability status
Simple GLP Basic view information
Last calibration date, oset, average slope
Full GLP Simple GLP information and calibration point details
Table Measurements updated every second are displayed
in table
Graph (Plot) Measurement versus time graph can be panned or zoomed (pinch-to-zoom technology)
Logging Type Automatic, Manual, Autohold
Number of records 50 000 maximum per le
Stores at least 1 000 000 data points per user
Automatic interval 1, 2, 5, 10, 30 seconds
1, 2, 5, 10, 15, 30, 60, 120, 150, 180 minutes
Sample ID Incremental mode
Export option .csv le format
Users Up to 9 users and the default administrator account
Connectivity USB-A 2 ports for keyboard input or USB thumb drive
  USB-C 1 port for PC connectivity and USB-C type thumb drive
Wi-Fi & Ethernet FTP
Web server   Log transfer and download Email
RS232 Connecting peripherals
Power supply DC adapter 100-240AC to 24VDC 2.5A
Environment 0 - 50 °C / 32 - 122 °F / 273 - 323 K maximum 95% RH non-condensing
Dimensions 205 x 160 x 77 mm (8.0 x 6.2 x 3.0 “)