Eppendorf BioSpectrometer® Basic

The Eppendorf BioSpectrometer® basic makes your work easier with the following features: simplest handling, guided software procedures, operation directly on the device, preprogrammed methods, automatic evaluation of measurement data and clear presentation of results.

Brand: Eppendorf

Learn more

Product CodeProduct Product Description Price (Excl gst) Qty Add
EP6135000033 Eppendorf BioSpectrometer® Basic Eppendorf BioSpectrometer® Basic Login for price  
EP6135000907 Eppendorf BioSpectrometer® Basic with Eppendorf µCuvette™ G1.0 Eppendorf BioSpectrometer® Basic with Eppendorf µCuvette™ G1.0 Login for price  

Go to checkout

The Eppendorf BioSpectrometer basic is a small, very compact spectrophotometer that combines easy handling with highly reproducible analysis. It can measure and record UV/Vis spectral ranges as well as measure individual wavelengths from 200 nm to 830 nm. Its freely selectable wavelengths provide maximum flexibility for all current and future applications. In addition to preprogrammed methods for standard measurements, users can create and store their own methods. Optimized menu navigation guides through the individual methods in a step-by-step process. All required entries are visible, no essential data is omitted. A help box explains the operational sequences and is available in 5 different languages.

Features

  • Absorbance measurement for one or more wavelengths, recording of wavelength scans from 200 nm to 830 nm (increment: 1 nm)
  • Automatic evaluation and storage of measurement data with clear presentation of results
  • Spectral graph to display sample purity with automatic ratio calculation
  • Preprogrammed applications for quick start
  • Freely programmable applications with evaluation via factor, standard or standard series and freely programmable parameters
  • Two-wavelength method with substraction or division evaluation
  • Integrated application and results memory
  • Data transfer via USB interface, Ethernet or via e-mail and direct printout of results
  • Guided software process to minimize errors
  • Integrated self-test and calibration history

 

Absorption measuring range

0 A – 3.0 A (260 nm)

BSA concentration range

7.5 ng/µL – 45.45 µg/µL

BSA concentration range (UV 280 nm) µCuvette G1.0

75 ng/µL – 45.45 µg/µL
±0.03 µg/µL at A = 0
±0.076 µg/µL at A = 1 (0.5 %)

BSA concentration range (UV 280 nm) UVette 10 mm

7.5 ng/µL – 4.5 µg/µL
±0.003 µg/µL at A = 0
±0.007 µg/µL at A = 1 (0.5 %)

BSA concentration range (UV 280 nm) UVette 2 mm

37.5 ng/µL – 22.5 µg/µL
±0.015 µg/µL at A = 0
±0.038 µg/µL at A = 1 (0.5 %)

Beam receiver absorption

CMOS photodiode array

Cuvette shaft

12.5 mm × 12.5 mm

Light path height

8.5 mm

Light source absorption

Xenon flash lamp

Measuring principle absorption

Single-beam absorption spectrophotometer with reference beam

Method memory

>100 method programs

Operator guidance language

Spanish, Italian, French, English, German

Power consumption

Approx. 15 W during operating step
Approx. 5 W with dimmed display

Random error absorption

≤0.002 at A = 0
≤0.005 (0.5 %) at A = 1

Random error wavelength

±1 nm

Smallest step size

1 nm

Spectral bandwidth

≤4 nm

Systematic error absorption

±1 % (A = 1)

Systematic error wavelength

≤0.5 nm

Wavelength range

200 nm – 830 nm

Wavelength range absorption

Scan (nm): 200 – 830 at 1 nm increments

dsDNA concentration range

2.5 ng/µL – 1,500 ng/µL

dsDNA concentration range (UV 260 nm) UVette 10 mm

2.5 ng/µL – 150 ng/µL
±0.1 ng/µL at A = 0
±0.25 ng/µL at A = 1 (0.5 %)

dsDNA concentration range (UV 260 nm) UVette 2 mm

12.5 ng/µL – 750 ng/µL
±0.5 ng/µL at A = 0
±1.25 ng/µL at A = 1 (0.5 %)

dsDNA concentration range (UV 260 nm) µCuvette G1.0

25 ng/µL – 1,500 ng/µL
±1 ng/µL at A = 0
±2.5 ng/µL at A = 1 (0.5 %)

Interfaces

› USB master: for USB stick and thermal printer DPU-S445 
› USB slave for connecting to a PC 
› Serial interface RS-232: for thermal printer DPU-414 
› Ethernet interface RJ45: For connecting to a network printer

Power supply

230 V, 50 – 60 Hz

Dimensions (W × D × H)

295 × 400 × 150 mm / 11.6 × 15.7 × 6 in

Weight w/o accessories

5.4 kg